SPECIFICATIONS


A. GAS SAMPLE CONDITIONS

Sample Inlet Pressure:

0 - 100 (psig)

Flow Rate:

Up to 500 sccm

Sample Gases:

Ammonia (NH3), Phosphine (PH3), Arsine ( AsH3), and Inert

Sample Line Temperature:

Up to 60°C

B. GAS FLOW RATE
Materials of Construction: 316L stainless steel (optional Hastelloy®)
Wetted Components: 10 Ra surface finish
Gas Connection: Sample inlets & outlets - 1/4" VCR™
Leak Tested to: <2 x 10-8 mbar • l /sec

C. ELECTRICAL
Alarm Indicators: User-programmable alarm set points

Power Requirements:

90-240 VAC 50/60 Hz

Power Consumption:

200 Watts max.

Output Signals:

 

      Recorder

Isolated 4-20mA or 0-20 mA output per sample inlet

      Alarm

Form-C relay per sample inlet
  Form-C relay for system fault
  (Laser status & ring-down status)

User Interface:

10.4" color VGA display with touch screen, PS2 Mouse
and Keyboard connection, 10BaseT Ethernet, 3.5" USB, RS-232

D. DIMENSIONS

Mounting:

14" x 19" x 29" (H x W x D)

Weight:

100 pounds

PERFORMANCE

Lowest Detection Limit:

8 ppb (in hydride), 0.5 ppb (in inert)

Sensitivity

5 ppb (in hydride), 0.3 ppb (in inert)

Accuracy (greater of)

4% of reading or +/- 5 ppb

Speed of Response (typical):

Response to 50 ppb intrusion - less than one minute
Operating Range: 0 - 8 ppm (in hydride), 0 - 5 ppm (in inert)
Environmental Conditions: 10° - 40° C
Storage Temperature: -10° - 50° C

TECHNOLOGY

Approvals:

CE: LVD & EMC approved

Method:

Cavity Ring-Down Spectroscopy
Patents: U.S. Patent # 5,528,040

 

Other patents pending